LIFEPAK® 15 - CO2 Input Cleaning Procedure

Modified on Mon, 30 Mar, 2020 at 4:38 PM

Q: How to clean the CO2 input in LIFEPAK® 15?


A: By following these steps:

  • Clean when readings do not appear when a new filter line is screwed into the CO2 input.
  • Clean the threads on the inside surface of the CO2 input using compressed air.
  • Used only on the outer two rings.
  • The inner ring and interior of the inner ring should not be cleaned.Errors and omission excepted

Was this article helpful?

That’s Great!

Thank you for your feedback

Sorry! We couldn't be helpful

Thank you for your feedback

Let us know how can we improve this article!

Select at least one of the reasons
CAPTCHA verification is required.

Feedback sent

We appreciate your effort and will try to fix the article