Q: How to replace the lid in LIFEPAK® CR2?


A: Follow these steps:

  • Remove the battery from the device
  • Open the lid.
  • Push in one end of lid hinge to release the lid from the case.
  • Insert one end of the new lid into the small hole and push in the other end of the lid and snap into the place
  • Close the lid.
  • Insert the battery.
  • Verify the readiness light is flashing in green every six seconds.