LIFEPAK® 15 - CO2 Input Cleaning Procedure

Modified on Mon, 30 Mar 2020 at 04:38 PM

Q: How to clean the CO2 input in LIFEPAK® 15?


A: By following these steps:

  • Clean when readings do not appear when a new filter line is screwed into the CO2 input.
  • Clean the threads on the inside surface of the CO2 input using compressed air.
  • Used only on the outer two rings.
  • The inner ring and interior of the inner ring should not be cleaned.Errors and omission excepted