LIFEPAK® CR2 - Replacing the Battery

Modified on Mon, 30 Mar 2020 at 04:51 PM

Q: What should I know about replacing the battery of LIFEPAK® CR2?


A: The following:

  • Device will automatically begin a self test when the battery is installed
  • Do not open the lid till you hear 3 audible tones
  • Audible tones will not sound if re-installing the same battery
  • Wait till the Readiness indicator is flashing every 6 seconds before opening the lid


Note: The battery must be installed within 1 year of this date to attain a 4-year service life. The maximum service life for the battery is 5 years after the date of manufacture or 4 years after the date installed in the defibrillator, whichever occurs first.